Edan Blood Gas and Chemistry Analyser Model i15
Edan i15 Blood Gas and Chemistry Analyser
The Edan i15 is a portable point-of-care blood gas and chemistry analyser delivering results within one minute of sample aspiration from a 140 µl whole blood sample. Combining blood gases, electrolytes, metabolites and haematocrit in a single multi-parameter single-use cartridge, the i15 eliminates the need for multiple separate analysers at the bedside. A 7-inch colour touchscreen, auto-sampling mechanism and built-in multimedia tutorials minimise training requirements and handling time. A 5000 mAh lithium-ion battery supports 50 continuous tests per charge, enabling use in emergency departments, intensive care units, wards, outpatient clinics and field settings without a mains connection. Up to 10,000 patient records are stored onboard, with 4 USB ports, LAN and Wi-Fi for flexible data management and optional EMR integration. A triple quality control guarantee system incorporates calibration, liquid controls and an external electronic simulator for verified result accuracy.
Key Features
- Results in 1 minute - from sample aspiration to reported result in under 60 seconds
- 140 µl sample volume - small whole blood sample from arterial line, capillary or syringe
- Multi-parameter single-use cartridge - blood gases, electrolytes, metabolites and haematocrit in one cartridge; no cross-contamination risk; room temperature storage with long shelf life
- Available cartridge configurations - BG3, BG4, BG8, BG9, BG10 and BC4 to match clinical need
- Zero maintenance - single-use cartridge system eliminates sensor cleaning, calibration fluid handling and maintenance procedures
- 7-inch colour LCD touchscreen - 800 x 480 resolution; intuitive interface with built-in multimedia tutorial videos
- Barcode and QR code scanner - acquire patient information without manual data entry
- 50 tests per charge - 5000 mAh rechargeable lithium-ion battery; portable use across clinical areas without mains power
- Triple QC guarantee - internal liquid calibration, 3 or 5 level liquid quality controls, and external electronic simulator for comprehensive result verification
- Up to 10,000 patient record storage - onboard memory with patient data management
- 4 x USB 2.0, RS232, LAN and Wi-Fi - flexible data export, EMR integration and optional data management software
- Built-in printer - onboard result printing without external hardware
- 3.65 kg portable design - lightweight and compact for bedside, ward and point-of-care use
Measured Parameters
| Parameter | Reportable Range | Resolution | Method |
|---|---|---|---|
| pH | 6.5 to 8.0 | 0.001 | Potentiometric sensor |
| pO2 | 10 to 700 mmHg | 0.1 | Amperometric sensor |
| pCO2 | 10 to 150 mmHg | 0.1 | Potentiometric sensor |
| Na | 100 to 180 mmol/L | 0.1 | Potentiometric sensor |
| K | 2.0 to 9.0 mmol/L | 0.01 | Potentiometric sensor |
| Ca | 0.25 to 2.5 mmol/L | 0.01 | Potentiometric sensor |
| Cl | 65 to 140 mmol/L | 0.1 | Potentiometric sensor |
| Glucose | 20 to 700 mg/dL / 1.1 to 38.9 mmol/L | 1 / 0.1 | Amperometric sensor |
| Lactate | 2.7 to 180.2 mg/dL / 0.3 to 20 mmol/L | 0.1 / 0.01 | Amperometric sensor |
| Haematocrit | 10 to 75% / 0.10 to 0.75 fraction | 1% / 0.01 | Conductance sensor |
Calculated Parameters
cH+, HCO3-act, HCO3-std, BE(ecf), BE(B), BB(B), ctCO2, sO2(est), Ca++(7.4), AnGap, tHb(est), pO2(A-a), pO2(a/A), RI, pO2/FIO2, and temperature-corrected values: cH+(T), pH(T), pCO2(T), pO2(T), pO2(A-a)(T), pO2(a/A)(T), RI(T), pO2(T)/FIO2, mOsm
Technical Specifications
| Specification | Detail |
|---|---|
| Manufacturer | Edan Instruments, Inc. |
| Model | i15 |
| Sample volume | 140 µl |
| Throughput | Results within 1 minute of sample aspiration |
| Display | 7-inch colour LCD, 800 x 480 |
| Input | Touchscreen and barcode/QR code scanner |
| Battery | 5000 mAh rechargeable lithium-ion |
| Battery performance | 50 samples continuous testing per charge |
| Power supply | 100 to 240V AC, 50/60Hz |
| Dimensions (W x H x D) | 238 x 310 x 153 mm |
| Weight | 3.65 kg |
| Operating temperature | 10°C to 31°C |
| Operating humidity | 25% to 80% RH |
| Operating pressure | 70 to 106.6 kPa |
| Patient record storage | Up to 10,000 |
| Quality control | 3 or 5 level liquid QC; external electronic simulator |
| Interface | 4 x USB 2.0, 1 x RS232, WLAN, LAN |
| Built-in printer | Yes |
Please note: Cartridge type must be selected according to the parameters required for each clinical application. Cartridges are single-use only and must be stored at room temperature within shelf-life dates. Not all cartridge configurations are available in all countries — confirm UK availability before ordering.