{"product_id":"blood-bank-refrigerator-labcold-rsbg1030md","title":"Labcold RSBG1030MD Blood Bank Refrigerator - 88ltr","description":"\u003c!-- ═══════════════════════════════════ SECTION 1 : DESCRIPTION, REFERENCES, FEATURES, NOTICES ═══════════════════════════════════ --\u003e\n\u003ch2\u003eLabcold RSBG1030MD Compact Blood Bank Refrigerator, 88 Litres, Glass Door, Pre-owned\u003c\/h2\u003e\n\u003cp\u003eThe RSBG1030MD is Labcold's compact blood bank refrigerator, 88 litres and holding around thirty 250 ml blood bags, running at 4 degrees Celsius with a seven day chart recorder and a full alarm set. It is a benchtop sized cabinet at 490 mm wide, which is what makes it the unit for a satellite fridge sited where blood is actually needed rather than only in the transfusion laboratory.\u003c\/p\u003e\n\u003cp\u003eThe difference between this and a pharmacy refrigerator is the way temperature is measured, and it is not a detail. A pharmacy or domestic cabinet reads the air inside the chamber, and air changes temperature the instant a door opens. A blood bank refrigerator reads a liquid load simulator as well: a sealed container of fluid chosen to match the thermal mass of a blood bag, so the figure on the display tracks what the blood is doing rather than what the air is doing. This cabinet shows both the chamber and the liquid load simulator on its touch panel, which is the measurement that actually matters when the door has just been opened twice in a minute.\u003c\/p\u003e\n\u003cp\u003eNHSBT specify that red cells are stored at 4 degrees Celsius plus or minus 2, in an approved blood refrigerator fitted with a temperature recorder and an alarm. This cabinet is built to that requirement: a seven day chart recorder, an adjustable audible alarm, a core temperature action alarm, remote alarm contacts for connection to a building management or on-call system, and a blood audit interface. Storage of blood for transfusion in the United Kingdom falls under the Blood Safety and Quality Regulations 2005, and the cabinet is certified as a medical device and built to BS 4376 Part 1.\u003c\/p\u003e\n\u003cp\u003eThe practical case for a small cabinet of this kind is the thirty minute rule. A unit of red cells that spends more than thirty minutes out of controlled storage cannot simply go back into the issue fridge, and beyond sixty minutes it is a quarantine question. A compact blood bank refrigerator close to theatre, to the emergency department or to a remote issue point keeps that clock short. The glass door with clear inner doors lets stock be counted without opening the chamber, which is itself a thermal event, and the keypad lock controls who gets into it.\u003c\/p\u003e\n\u003ch3\u003eReferences\u003c\/h3\u003e\n\u003ctable\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eModel\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eRSBG1030MD\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eBrand\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eLabcold\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eProduct\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eCompact blood bank refrigerator\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eCapacity\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e88 litres\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eBlood bag capacity\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eAround 30 bags at 250 ml\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDoor\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eGlass, with clear inner doors\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eSolid door equivalent in the Labcold range\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eRSBB prefix, a separate model\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eOperating temperature\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e4 degrees Celsius\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRecorder\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eSeven day chart recorder\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eCondition\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003ePre-owned\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003ch3\u003eFeatures\u003c\/h3\u003e\n\u003cul\u003e\n\u003cli\u003eLiquid load simulator alongside the chamber probe, so the displayed temperature tracks the blood rather than the air.\u003c\/li\u003e\n\u003cli\u003eTouch panel showing chamber and liquid load simulator temperatures together.\u003c\/li\u003e\n\u003cli\u003eSeven day chart recorder, giving the temperature record NHSBT call for.\u003c\/li\u003e\n\u003cli\u003eAdjustable audible alarm, with symbols on the panel showing normal or alarm status at a glance.\u003c\/li\u003e\n\u003cli\u003eCore temperature action alarm.\u003c\/li\u003e\n\u003cli\u003eRemote alarm contacts, for connection to a building management system or an on-call alarm route.\u003c\/li\u003e\n\u003cli\u003eBlood audit interface.\u003c\/li\u003e\n\u003cli\u003eKeypad lockable door, for controlled access to transfusion stock.\u003c\/li\u003e\n\u003cli\u003eGlass door with clear inner doors, so stock is counted without opening the chamber.\u003c\/li\u003e\n\u003cli\u003eTwo adjustable shelves.\u003c\/li\u003e\n\u003cli\u003eStainless steel interior over a powder coated steel exterior.\u003c\/li\u003e\n\u003cli\u003eBenchtop footprint at 490 mm wide, for siting close to the point of use.\u003c\/li\u003e\n\u003cli\u003eCertified as a medical device and built to BS 4376 Part 1.\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003cdiv style=\"background: #fff3cd; border-left: 5px solid #ffc107; padding: 16px 20px; margin: 0 0 22px; border-radius: 4px;\"\u003e\n\u003cp\u003e\u003cstrong style=\"color: #856404;\"\u003ePoints to note before ordering\u003c\/strong\u003e\u003c\/p\u003e\n\u003cul\u003e\n\u003cli\u003e\n\u003cstrong\u003eGlass door model.\u003c\/strong\u003e The RSBG prefix is the glass door cabinet. The solid door version of this range carries the RSBB prefix and is a different model, so paperwork quoting RSBB does not describe this unit.\u003c\/li\u003e\n\u003cli\u003e\n\u003cstrong\u003eCapacity depends on what is stored.\u003c\/strong\u003e The figure of around thirty bags assumes 250 ml units on the shelves supplied. Larger units, overbagged stock or a basket arrangement will reduce it.\u003c\/li\u003e\n\u003cli\u003e\n\u003cstrong\u003ePre-owned condition.\u003c\/strong\u003e Marks to the cabinet consistent with service use. Door seals, hinges, the lock and the recorder are checked and their condition stated in writing, and Elarix Medical photographs the actual unit on request. Decontamination to local policy before use.\u003cbr\u003e\n\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003c\/div\u003e\n\u003c!-- ═══════════════════════════════════ SECTION 2 : TECHNICAL SPECIFICATION ═══════════════════════════════════ --\u003e","brand":"Labcold","offers":[{"title":"VAT Exempt","offer_id":59937092501829,"sku":"LAB4190V","price":349.99,"currency_code":"GBP","in_stock":true},{"title":"VAT Included","offer_id":59937092534597,"sku":"LAB4190V","price":349.99,"currency_code":"GBP","in_stock":true}],"thumbnail_url":"\/\/cdn.shopify.com\/s\/files\/1\/0805\/2716\/3717\/files\/Labcold_RSBG1030MD_Compact_Blood_Bank_Refrigerator.jpg?v=1793121492","url":"https:\/\/selfimed.com\/ro\/products\/blood-bank-refrigerator-labcold-rsbg1030md","provider":"SelfiMed UK","version":"1.0","type":"link"}