Hitachi
Hitachi Hi-Vision Avius Ultrasound Machine System
Hitachi HI VISION Avius Diagnostic Ultrasound Machine System
The Hitachi HI VISION Avius Ultrasound Machine is a high performance diagnostic ultrasound platform designed for accuracy, speed and clinical efficiency. With advanced image processing and an ergonomic user interface, this ultrasound system supports a wide range of specialties including radiology, endocrinology, neurology, rheumatology, gastroenterology and urology.
It delivers strong penetration with high temporal, spatial and contrast resolution for dependable diagnostic results.
This ultrasound machine features customised imaging presets and fast image processing to support confident clinical decision making. The intuitive operator console provides smooth control of imaging parameters, while technologies such as harmonic imaging, real time tissue elastography, 4D capability and patient scan selection enhance workflow. The system is compatible with multiple probe types including cardiac, convex, linear, laparoscopic and endocavitary, providing broad clinical versatility.
Key Features
- Sony Video Graphic Printer UP-895CE
- Footswitch included
- Software version V03-61 Step 3.5
- Customised imaging presets
- High speed image processing
- Real time tissue elastography
- Harmonic imaging with dynamic contrast
- 4D imaging capability
- Advanced diagnostic imaging tools
- Patient scan selector
Specifications
- Brand: Hitachi
- Applications: Endocrinology, Radiology, Neurology, Rheumatology, Urology, Gastroenterology
- Weight: 125 kg
- Size: 45 × 70 × 135 cm
- Probe Types: Cardiac, Convex, Linear, Laparoscopic, Endocavitary (probes not included)
Engineer Report
The HI VISION Avius ultrasound system powers up and completes initialisation successfully. The LCD monitor, control panel, gain controls and trackball are fully operational. Patient registration and image storage functions work correctly. Functional modes tested include B mode, CFI, PW and M mode.
Testing with a Pentax EG-3670URK EUS radial probe (not included) produced fair image quality. The flexible videoscope was tested on a compatible stack system and displayed very good image quality. Foot control and printer are functioning correctly. The system is in good physical condition with minor cosmetic scuffs. Both ultrasound and videoscope functions operate normally.
Enabled Options
- EZU-FC11W DICOM Worklist
- EZU-FC11P DICOM Printer
- EZU-DVA Digital Archiving
- EZU-NW1 Network Interface