Philips Respironics BiPAP Machine A40 Unit with 1096770 Detachable Battery Module with Li-ion Battery
£599.99
This item
Philips DreamWisp Nasal CPAP Mask Fit Pack
£84.00
Philips DreamStation CPAP Machines Travel Bag
£24.99
This pre-owned BiPAP machine is a Philips Respironics BiPAP A40 Pro bi-level non-invasive ventilator built for patients with chronic and complex respiratory insufficiency. It is indicated for adult and paediatric patients above 10 kg and is used in hospitals, sleep laboratories and home ventilation programmes. This unit is supplied with the 1096770 detachable battery module and lithium-ion battery, together with the heated humidifier and water chamber.
The A40 Pro adapts therapy to changing patient conditions and delivers the lowest effective pressure while maintaining ventilation quality. AVAPS holds a target tidal volume by adjusting pressure support, AVAPS-AE additionally monitors upper airway resistance and adjusts EPAP to keep the airway patent, and Automated Airway Management supports patients with overlapping breathing syndromes. Auto-Trak synchronisation with leak compensation keeps the device responsive to weak respiratory drive, which matters for neuromuscular and hypoventilation cases.
Visual Leak gives a real-time indication of total circuit leak, which helps the clinical team resolve mask fit problems quickly. The device connects to Philips Care Orchestrator for remote therapy review and patient management. Operation is quiet at around 30 dBA and the housing is compact enough for long-term domiciliary use while retaining the performance expected in a hospital NIV service.
| Item | Detail | Condition |
|---|---|---|
| BiPAP A40 Pro unit | Philips Respironics bi-level non-invasive ventilator, main device | USED |
| Heated humidifier | Attached heated humidifier unit for the A40 platform | USED |
| Water chamber | Humidifier water chamber, cleaned and inspected | USED |
| Battery module | 1096770 detachable battery module | USED |
| Lithium-ion battery | Battery cell for the detachable module | USED |
| Power supply | 24 V DC power supply with mains lead | USED |
| Item | Detail | Status |
|---|---|---|
| Mask and headgear | Any interface must be selected for the patient, for example the BMC F5A full face CPAP mask | SOLD SEPARATELY |
| Patient circuit | A fresh circuit is mandatory before use on a patient | SOLD SEPARATELY |
| Air inlet filters | Reusable foam filter and disposable ultrafine filter | SOLD SEPARATELY |
| Heated tube | Optional heated circuit tubing for the humidifier | SOLD SEPARATELY |
| SD card | Therapy data card | SOLD SEPARATELY |
| Modem and Care Orchestrator access | Connectivity hardware and any platform subscription | SOLD SEPARATELY |
| Oxygen tubing and source | Supplementary oxygen equipment where prescribed | SOLD SEPARATELY |
| Carry case | Transport bag for the device | SOLD SEPARATELY |
| Distilled water | Required for the humidifier chamber | SOLD SEPARATELY |
Mandatory before any clinical use
| Manufacturer | Philips Respironics |
| Model | BiPAP A40 Pro |
| Device type | Bi-level non-invasive ventilator, non-life support |
| Patient group | Adult and paediatric, above 10 kg body weight |
| Pressure range | 4 to 40 cm H2O |
| Ventilation modes | CPAP, S, S/T, PC, T, AVAPS-AE |
| Therapy features | AVAPS, AVAPS-AE, Automated Airway Management, Visual Leak |
| Trigger type | Auto-Trak, sensitive Auto-Trak, flow trigger |
| Breath rate | 0 to 40 breaths per minute in S/T and PC, 4 to 40 breaths per minute in T |
| Inspiratory time | 0.5 to 3.0 seconds |
| Humidification | System One heated humidifier, heated tube compatible |
| Supplementary oxygen | Low flow, up to 15 L/min at 10 psi |
| Settable alarms | Circuit disconnect, apnoea, low tidal volume, low minute ventilation, high respiratory rate, low SpO2 |
| Sound level | Approximately 30 dBA at 10 cm H2O |
| Weight | Approximately 2 kg, device only |
| Dimensions, W x L x H | 22.23 x 18.42 x 10.80 cm |
| AC power | 100 to 240 V AC, 50/60 Hz |
| DC power | 24 V DC power supply |
| Battery module | 1096770 detachable battery module with lithium-ion battery |
| Connectivity | Philips Care Orchestrator compatible |
| Condition | Pre-owned, cleaned and inspected |
Pre-owned battery module
The lithium-ion battery supplied with this unit is pre-owned. The usable capacity of any lithium-ion cell falls with age and cycle count, so the runtime achieved will be lower than that of a fresh cell and cannot be stated for this unit. The battery must be assessed by the commissioning engineer and must not be relied upon as the sole power source for a patient. Where mains independence is clinically required, a fresh battery should be fitted before the device enters service.
Pre-owned electrical equipment
The unit has been cleaned and visually inspected. An engineer electrical safety test is recommended before clinical use, alongside the ventilator function test described above. The photographs form part of the description and show the actual condition of the item offered.
Supplementary oxygen safety
Where oxygen is added to the circuit, keep the device and circuit away from smoking, naked flames and any source of ignition. Do not use oil, grease or petroleum based products on or near the equipment. Use the device in a ventilated room and keep the air intake and filter area unobstructed. Oxygen flow must be set by the responsible clinician and the oxygen source must be turned off when the device is not running.
Humidifier and water chamber
Fill the chamber with distilled water only. Tap water leaves mineral deposits that damage the chamber and reduce humidifier performance. Empty and dry the chamber daily and clean it before first use in a new patient. The chamber is a consumable component and should be renewed at the interval given in the instructions for use.
| Component | Action | Interval |
|---|---|---|
| Reusable foam filter | Wash in warm water with mild detergent, air dry fully before refitting | Weekly |
| Disposable ultrafine filter | Fit a fresh filter, or sooner if visibly soiled | Monthly |
| Patient circuit | Clean per the circuit instructions, inspect for splits and leaks | Weekly |
| Water chamber | Empty, rinse and dry, refill with fresh distilled water | Daily |
| Device exterior | Wipe with a damp cloth, device disconnected from the mains | Weekly |
| Mask and headgear | Clean per the interface instructions | Per interface guidance |
Intervals above are general guidance for a single patient in a domiciliary setting. Institutional decontamination policy and the manufacturer instructions for use take precedence.
Ordered before 1pm Monday to Friday, dispatched the same working day from Leeds. Deliveries run Monday to Saturday, never Sunday.
UK mainland, light items: 1 to 2 working days. Heavy and bulky: 2 to 4 working days.
International orders ship on a Delivered At Place basis. Import duty, local tax and customs clearance fees are charged by the destination country and are the buyer's responsibility.
Unopened items can be returned within 30 days of delivery. Tell us before you send anything back so we can give you a returns reference.
Opened items that touch the airway cannot be returned for hygiene reasons unless they are faulty.
This does not affect your statutory rights.
Refunds are issued to the original payment method once the returned item reaches us and has been checked, normally within five working days.
Faulty items are refunded in full, including the delivery you paid.
Information on this page is provided for guidance and does not replace advice from a qualified clinician. Always read the manufacturer's instructions before use, and check that a device is suitable for the patient and the prescribed therapy.
Goes on the same order.
More in the same category.
Picking up where you left off.
Meical Equipment Advice and Guidance
August 26, 2026
Avoid the most common fitting, cleaning and ordering mistakes with the ResMed QuietAir Elbow. Practical, factual guidance for clinicians ...
Read Article
August 26, 2026
A plain factual guide to the Sidhil Innov8 iQ Electric profiling bed: what it does, what to check before ordering and what maintenance in...
Read Article
August 26, 2026
Avoid the most common installation, maintenance and ordering mistakes with the ALVERA PureGuard™ Universal In-Line filter, with plain fac...
Read Article