The Oxford Midi 180 Patient Hoist is a mobile, electrically powered lifting device designed to transfer patients who cannot weight-bear independently. Used widely across wards, care homes and community settings, it supports safe manual handling practice when operated within its specified parameters. This guide covers what matters most: setup, daily use, compliance obligations and routine care.
The Oxford Midi 180 is rated to a safe working load of 180 kg. That figure is the ceiling for the complete system, meaning the hoist, the sling and all attachments combined. If a sling with a lower working load is used, the sling's rating governs. Never exceed the lowest safe working load present anywhere in the assembled system.
The hoist is intended for non-weight-bearing transfers: bed to chair, chair to toilet, floor recovery and similar tasks. It is not a stand aid. Using it for a task outside its designed scope introduces risk and may void the manufacturer warranty.
An electric actuator raises and lowers the boom. Leg-spread is controlled manually or electrically depending on the variant. Battery condition directly affects lift performance. A depleted battery may stall mid-lift, which is a safety-critical event. Check the charge indicator before every session.
The Oxford Midi 180 uses a spreader bar designed for Oxford-pattern loop attachments. Confirm sling compatibility with the sling manufacturer before use. Slings are consumable items and are always sold separately. Inspect every sling before each use for fraying, torn stitching or deformed loops. A sling that shows any of these signs must be withdrawn from service immediately.
![]() |
Oxford Midi 180 Patient Hoist New £599.99 Buy nowView product |
The Lifting Operations and Lifting Equipment Regulations 1998 (LOLER) require a thorough examination of the Oxford Midi 180 by a competent person at intervals not exceeding six months. This applies to every sling used with it as well. The thorough examination is not an in-house visual check. It must be carried out by a person with the specific knowledge, training and authority to make a written report.
Responsibility for arranging the examination rests with the operating organisation, not with the supplier. A newly received hoist must be examined before it enters service unless a valid, in-date thorough examination report transfers with it.
The Provision and Use of Work Equipment Regulations 1998 (PUWER) require the hoist to be maintained in a safe condition and that maintenance records are kept. Planned preventive maintenance, electrical safety testing and actuator checks are the responsibility of the operating organisation's competent maintenance function. These are not included in a standard pre-owned supply unless the listing explicitly states otherwise.
Keep the following on file for every hoist in service:
A clinical or care team member should run a brief pre-use check at the start of every shift the hoist will be used. This is distinct from the LOLER thorough examination and does not replace it.
The Oxford Midi 180 requires clear floor space to manoeuvre safely. Minimum doorway width and bed clearance requirements should be confirmed against the manufacturer's dimensional data before the hoist is commissioned in a new space. Attempting a transfer in a cluttered or confined space increases risk to the patient and to the handler.
Use only the charger specified for the Oxford Midi 180. Connecting an incompatible charger may damage the battery management system. Charge after every session or at the end of each shift rather than waiting for the battery to be fully depleted. If the hoist is to be stored unused for an extended period, follow the manufacturer guidance on storage charge state to avoid deep discharge.
The Oxford Midi 180 is classified as a non-critical device that contacts intact skin via a sling interface. It must be decontaminated between patients according to the operating organisation's infection prevention and control protocol. Use only cleaning agents compatible with painted steel and plastic surfaces. Refer to the manufacturer's decontamination guidance; that guidance takes precedence over any general recommendation in this article.
The table below outlines a typical care schedule. Adapt it to your organisation's specific maintenance policy and the manufacturer instructions.
Slings are consumable items. They are not reusable indefinitely and must be inspected and retired according to the sling manufacturer's guidance. Batteries are time-limited components. A used hoist supplied without a battery, or with a battery of unknown service history, should have the battery assessed or replaced before the hoist enters clinical service. Chargers, slings and spreader bars are typically sold separately. Confirm what is included with any specific listing before ordering.
The Oxford Midi 180 Patient Hoist is a well-established mobile lifting device suited to a broad range of non-weight-bearing transfer tasks up to 180 kg. Getting the most from it requires attention to four areas: compliance with LOLER and PUWER, consistent pre-use checks, correct charging and decontamination practice, and accurate record keeping. Slings are always a separate consideration and must be matched to the hoist and to the patient. The operating organisation carries responsibility for ongoing examination, maintenance and record keeping. Manufacturer instructions take precedence over general guidance in every case.
![]() |
Ready to move forward with the Oxford Midi 180 Patient Hoist New? Shop the Oxford Midi 180 Patient Hoist New |
SelfiMed UK
BLACK FRIDAY
+ up to −80% on selected items, no code needed
Join our list for deal alerts, restocks and CPAP tips. Your code appears straight away.
Free UK mainland delivery over £300 · next-day dispatch before 1pm
By signing up you agree to receive marketing emails from SelfiMed UK. Unsubscribe at any time. Code valid on all products until Monday 30 November 23:59. One discount per item: items already at −50% or −80% keep that deal instead.
It's already in your basket, so 10% comes off at checkout.
Valid on all products. One discount per item. Ends Monday 30 November 23:59.
Lascia un commento